Nurse call system E-105 Cornell

Nurse call system E-105 Cornell
E-105

for Price / Quote
for Documentation
for Questions

Description: Cornell Model E-105 Stations are typically used for initiation of calls of any desired priority. May be used as a nurse follower or emergency call station. Operation: The On/Off toggle switch is used to place and cancel calls and a red LED indicates call. Mounting: The satin finish stainless steel faceplate with silk-screened designations may be flush mounted on a standard outlet box with single-gang plaster ring, or surface mounted on a Wiremold 5748 or equivalent. Engineering Specifications: The contractor shall furnish and install Cornell Model E-105 stations at the noted locations for use as emergency call or nurse follower stations. Faceplates shall be satin-finish stainless steel with silk-screened call placed, on, and off designations clearly marked. Call indicator shall be red light emitting diode. Incandescent lamps are not acceptable. On-off switch shall be toggle. For flush mounting, back box shall be standard outlet box with single-gang plaster ring; for surface mounting, Wiremold 5748 or equivalent. Technical Information: Power Requirements: 24VDC Operating Environment: 50-120°F (0-60°C) Indoor Non-condensing Physical Faceplate Dimensions: 4 ½ H x 2 ¾ W (1 Gang) Mounting: Single-Gang back box with 1¾ minimum depth Wiring: Minimum 22AWG
7915 North 81st Street,
Milwaukee,
WI 53223
× × × × × × ×