White blood cell cell counter / differential / digital / bench-top HemoCue® WBC DIFF HemoCue

White blood cell cell counter / differential / digital / bench-top HemoCue® WBC DIFF HemoCue
HemoCue® WBC DIFF

for Price / Quote
for Documentation
for Questions

HemoCue® WBC DIFF is a differential white blood cell counter which is commonly used for sample materials like capillary or venous whole blood and common parameters: Total Leukocytes (White Blood Cells) and Differential (in absolute and %) for: Neutrophils, Lymphocyter, Monocyter, Basophils and Eosinophild. For optimal performance of the HemoCue WBC DIFF system, the operating temperature should be: venous/capillary sample in EDTA tube: 18 - 30° C, <90% non-condensing humidity. Capillary sample from fingerstick: 18 - 25 °C, <90% non-condensing humidity. The analyzer can be stored at temperature 4-50 °C. Results: in about 5 minutes Sample volume: 10 µL. Weight: 1300g with batteries installed Power: AC Adapter or batteries Interface: Printer, keyboard, barcode reader and PC Connectivity: POCT1-A over Ethernet connection
  • Display type:digital
  • Configuration:bench-top
  • Cell type:white blood cell
ulica Zapolskiej 7,
Katowice
× × × × × × ×